Auscultate (listen with a stethoscope) over the precordial/epigastric (heart/upper stomach) area for a motorized “hum” and simultaneously visualize the controller for a green light or lit screen
Perform CPR only when there are no signs of flow or perfusion (the person is unresponsive, pulseless, or there is no evidence of the pump functioning [eg: no motor “hum”])
Perform a secondary assessment and treat per appropriate protocol
Notify the receiving facility promptly and consider early consultation with the VAD coordinator¹ or medical control, regardless of the patient’s complaint
Assure the power unit, extra batteries, and backup controller accompany patient
A trained support member² should remain with patient
EMT and up
Unless otherwise directed by medical control, transport patient to a facility capable of managing VAD patients
Medical control
Termination of resuscitation
Consultation with a VAD program provider
Key points
1 Institution VAD coordinator phone number may be found along with pump model information on a tag located on the pocket controller. Patients may also have a medical bracelet, necklace, or wallet card with this information.
2 Trained support members include family and caregivers who have extensive knowledge of the device, its function, and its battery units. They may act as a resource to the EMS provider when caring for a VAD patient.
One set of fully charged batteries provides 8-10 hours of power:
If the battery or power is low, the batteries need to be replaced immediately
Assist with the replacement of batteries if directed by patient/caregiver
Never disconnect both batteries at once as this can cause complete loss of VAD power
Keep the device components dry
The most common complication in VAD patients is infection. VAD patients are susceptible to systemic illness, sepsis, and septic shock due to their abdominal driveline as a conduit of infection.
Patients with a VAD are highly preload dependent and afterload sensitive. Low flow alarms are frequently due to MAP >90 mmHg. The devices are sensitive to alterations in volume status and careful volume resuscitation is often necessary.
VAD patients are heavily anticoagulated and susceptible to bleeding complications