Evaluation and/or transport of a patient with a total artificial heart (TAH)
All levels
Assess airway and breathing. Hypertension or volume overload can quickly cause pulmonary edema to develop.
Do not use an AED or cardiac monitor
Assess pulse, perfusion, and artificial heart function
If no pulse present:
Consider early consult with TAH coordinator¹ or medical control
Check for severed or kinked TAH driveline (address as possible)
Check battery position and power status (replace as possible)
Use the backup driver, or hand pump, if available
Do not perform chest compressions or place an AED
Assess blood pressure: goal blood pressure is >90 mmHg and <150 mmHg
Perform a secondary assessment and treat per protocol
If unresponsive with a pulse, evaluate for noncardiac etiologies
Notify the receiving hospital that your patient has a TAH while on scene or promptly after initiation of transport regardless of patient’s complaint
Assure that patient has drivers (compressor), hand pump, all batteries, and power cords for transport
Any trained support member² should remain with patient
Medical control
Termination of resuscitation
Consultation with a TAH program provider
Key points
1 Institution TAH coordinator phone number may be found along with pump model information on a tag located with the controller. Patients may also have a medical bracelet, necklace, or wallet card with this information.
2 Trained support members include family and caregivers who have extensive knowledge of the device, its function, and its battery units. They may act as a resource to the EMS provider when caring for a VAD patient.
TAH patients have had their heart removed and replaced with a rigid device which pneumatically pumps blood throughout the body. As these patients do not have a heart, there is no indication for an ECG or cardiac monitoring. A functioning TAH will not result in any measurable electrical activity.
TAH patients are on multi-agent anticoagulation and may have significant bleeding with minor injuries
The TAH patient has normal pulse and blood pressure detectable by conventional methods and are highly preload and afterload sensitive:
Target Blood Pressure is <150 mmHg and >90 mmHg
Pulse rate is set and regular, between 120-135 bpm