Automatic Transport Ventilator‡

Applies to adult and pediatric patients

  • This is a general resource document on the use of automatic transport ventilators, not a protocol. It is intended only for those who are separately equipped and trained. This does not supersede device-specific practice guidelines provided through agency education.

General Parameters

  • FiO2: Maintain SaO2 ≥92%
  • PEEP: 5 cm H2O (increase up to 10 cm H2O as needed to improve oxygenation)
  • Mode: A/C or SIMV
  • Pressure Support: 5 – 10 cm H2O, if in SIMV (if available)
  • Volume Control: Tidal volume (Vt) 6 – 8 mL/kg ideal body weight (maintain plateau pressure [Pplat] <30 cm H2O or PIP <35 cm H2O)
  • Rate: Child: 16 – 20 breaths/min; Adult: 12 – 14 breaths/min
  • I-Time: Child:0.7 – 0.8 seconds; Adult:0.8 – 1.2 seconds
  • Refer to the manufacturer’s ventilator operation manual for specific directions on how to operate your ventilator

Recommended Minimum Requirements for Automated Ventilator

  • Pressure limit / safety relief at a maximum of 40 cm H2O
  • Ability to adjust volume to 4-8 mL/kg ideal body weight
  • Ability to adjust rate in the minimum range of 10-30 breaths/min
  • Ability to add PEEP or PEEP valve in the minimum range of 5 - 10 cm H2O
  • Ability for patient triggered breaths (complete control ventilation is prohibited)

Initiating Mechanical Volume Ventilation

  • Use EtCO2 detection and pulse oximetry to evaluate the effectiveness of the ventilation technique and to verify artificial airway patency and position
  • Prepare the bag-valve-mask (BVM) for emergent use in case of a ventilator failure
  • Assure a secondary oxygen source with a minimum of 1000 psi in a D tank
  • Attach a ventilator to appropriate oxygen/air source
  • Attach a disposable ventilator circuit to ventilator
  • Attach a gas outlet, pressure transducer, and exhalation valve tubes to corresponding connectors
  • Select the appropriate mode, if applicable
  • Select the appropriate respiratory rate (RR). Titrate to appropriate EtCO2.
  • Adult: 12 – 14 breaths/min
  • Child: 16 – 20 breaths/min
  • Select the appropriate tidal volume (Vt) of 6 – 8 mL/kg ideal body weight
  • Select the appropriate inspiratory time (It), if applicable
  • Select the desired FiO2 if applicable. An FiO2 of 1.0 (100% O2) is a standard start and then should be titrated down to maintain SpO2≥92%.
  • Verify a high pressure alarm no higher than 40 cm H2O
  • Set PEEP to 5 cm H2O
  • Observe the delivery of several breaths
  • Evaluate the patient for adequate chest rise, ETCO2 and SpO2
  • Adjust the ventilator settings, as necessary, to improve clinical parameters
  • Record all set parameters on the patient transport record
  • Monitor and record PIP, if applicable

Key points

  • If at any time the ventilator should fail, or an alarm is received that cannot be corrected, the patient should be immediately ventilated with a bag-valve-mask (BVM) attached to a 100% oxygen source